Contaminated Sterilized Cannulated Instruments May Be Associated With Infections After Arthroscopic Surgery: A Systematic Review

Arthroscopy The Journal of Arthroscopic and Related Surgery · 10 Nov 2026

Rigour64/100
Clinical62/100
Priority56/100

Priority is the editorial-priority score. All scores are out of 100. Rigour is how well the study was done for its own design: 50 means nothing is known either way, and the design itself is not counted. Clinical is how far this paper should move patient care if its methods hold up: the most its design can claim, times the relevance of its question, times how clearly it answered. The size of a result is read only against a threshold the paper states itself.

Why it matters

Septic arthritis after arthroscopy is rare but serious, and inadequate reprocessing of hollow instruments is a modifiable cause that surgeons should consider when an infection cluster appears.

The evidence

This systematic review gathered 14 studies on sterilizing cannulated (hollow) orthopaedic instruments: 6 laboratory studies and 8 case series describing clusters of joint infections after arthroscopy, mostly ACL reconstruction. Lab work repeatedly found residual protein, biofilm, or viable bacteria inside lumens after cleaning and autoclaving (for example, 16 of 118 cannulated instruments that passed visual inspection were inadequately cleaned on ATP testing), and outbreak investigations rated contaminated cannulated instruments as a highly probable source in 3 of 8 clusters. There was no comparison group or pooled effect estimate, so the link to infection is an association rather than proven cause.

A coherent primary effect estimate is unavailable. Clinical relevance does not estimate treatment benefit.

Why this rigour score?

Systematic reviewNeutral starting point: 50, the same for every design

64/ 100

Raised the score

  • Prospectively registered+13.6
  • Comprehensive literature search+9.1

Lowered the score

  • No assessment of publication bias across 14 studies−18.2

Score capped at 64. A critical methodological failure for a review.

Room left by what was not reported. 7 of 8 criteria that apply to this design could be assessed from the paper. Had the others been reported, rigour could sit anywhere from 52 to 64. This is not a confidence interval.

Methods at a glance

Recorded assessment
  • Comprehensive literature searchMet
  • Risk of bias assessedMet
  • Bias considered in synthesisMet
  • Publication bias assessedNot met
  • Study exclusions justifiedMet
  • Prospective registrationMet

Not reported does not mean not done. Checks have different weights; this is not a count of boxes ticked.

See the full score breakdown
Neutral starting point (every design)
50
Prospectively registered
+13.6
Comprehensive literature search
+9.1
Risk of bias assessed in the included studies
+9.1
Risk of bias carried into the synthesis
+6.8
Excluded studies listed and justified
+4.5
No assessment of publication bias across 14 studies
−18.2
Methodological cap
−11
Final rigour score rounded
64 / 100

Rigour is how well the study was done for its own design. Every design starts at 50, which means nothing is known either way; 100 means every safeguard that applies to this design was shown. The points above are already scaled by how much this design can show: a check counts for more in a design with few applicable checks and for less in one with many. The design itself is not counted here: it sets the ceiling of the clinical score instead. Points are shown to one decimal place; the final score uses unrounded values.

Checklist from the stored appraisal. “Not met” describes the recorded criterion; it does not always produce a deduction. This overview supports reading the full study.

How the clinical score was reached

The question: 7/9. Three judgements of how much it matters to patients, which decision it informs and how closely it fits their care.

Patient consequence: moderate (2/3). A patient-experienced outcome of real but smaller or shorter-lived importance: early pain or function, speed of recovery, return to sport or work, minor complications, satisfaction, or a validated score at a short time point.

Decision relevance: moderate (2/3). Bears on a real decision, but indirectly: an association or risk factor that informs counselling or surveillance, a subgroup or secondary analysis, one component of a pathway, or a comparison that leaves out a main alternative.

Applicability: moderate (3/3). A recognisable practice population, but with a notable restriction: narrow eligibility, a single expert surgeon or centre, or an unusual protocol that would need adapting. Applicability only subtracts for a real mismatch, so moderate counts in full.

What it found: Points one way (×0.86). A direction is shown but not established: a verdict without a usable interval, a result from a trial the paper calls underpowered, a positive result from a small or fragile trial, one of many comparisons with no correction, or a review whose certainty is low or unrated. No primary interval or verdict was extracted, so this answer is the appraisal's own judgement of the paper, held to the same precision and certainty rules.

The design ceiling: 92. The most a systematic review can score here, however relevant its question and clear its answer. The design is counted here and nowhere else; rigour does not include it.

Clinical = 92 × 7/9 × 0.86 = 62.

For journal club

1.If two septic knees occur after ACL reconstructions in the same month at your hospital, what steps would you take in instrument review, staff cultures, and case scheduling before the next ACL case?

2.Should cannulated reamers and tendon harvesters be single-use or undergo extended cleaning and autoclaving protocols, and how would you weigh cost against the low baseline infection risk when counselling a hospital committee?

Original publication title

Contaminated Sterilized Cannulated Instruments May Be Associated With Infections After Arthroscopic Surgery: A Systematic Review

Read original paper